Abstract

A single quartz crystal microbalance (QCM) is cooled to 18 K to measure condensation rates inside of a retractable shroud enclosure. The shroud is designed to minimize condensate on fusion targets to be fielded at the National Ignition Facility (NIF). The shroud has a double-walled construction with an inner wall that may be cooled to 75 to 100 K.The QCM and the shroud system were mounted in a vacuum chamber and cooled using a cryocooler. Condensation rates were measured at various vacuum levels and compositions and with the shroud open or closed. A technique for measuring total condensate during the cooldown of the system with an accuracy of >1 × 10-6 g/cm2 was also demonstrated. The technique involves a separate measurement of the condensate-free crystal frequency as a function of temperature that is compared to the measurement for the cooldown trend of interest. The shroud significantly reduces the condensation rates of all gases and effectively eliminates H2O condensation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.