Abstract

Liquid argon time projection chambers (TPC) are widely used in neutrino oscillation and dark matter experiments. Detection of scintillation light in liquid argon TPC’s is challenging because of its short wavelength, in the VUV range, and the cryogenic temperatures (∼86 K) at which the sensors must operate. Wavelength shifters (WLS) are typically needed to take advantage of the high Photon Detection Efficiency (PDE) in the visible range of most of photondetectors. The Hamamatsu VUV4 S13370–6075CN SiPMs can directly detect VUV light without the use of WLS, which main benefit is an improved PDE at these short wavelengths, but also the visible light from WLS. The manufacturer (Hamamatsu Photonics K.K.) provides a complete characterization of these devices at room temperature; however, previous studies have indicated a decrease of the PDE at cryogenic temperature for VUV light. In this work, we present the measurement of the PDE of VUV4 SiPMs at cryogenic temperature for different wavelengths in the range [270, 570] nm. A dedicated measurement at 127 nm is also shown.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.