Abstract

The authors propose a new technique for detecting micro-electroretinograms (muERG) which uses light-emitting diode (LED) stimulation and frequency analysis. The advantage of this method is the applicability of linear system analysis to a muERG obtained by partial stimulation of a computer-controlled waveform. The paper discusses two basic techniques for detecting the muERG. One is Fourier analysis of a synchronised ERG obtained by sine-wave stimulus, and the other is correlation analysis using a random stimulus. The results of the two methods are compared, and the usefulness of the proposed method is indicated; the muERG is analysed by means of a model consisting of five components of a second-order transfer function with delay. Functions of the central and peripheral areas of the retina could be analysed in more detail by this technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.