Abstract

In this study the cumulative scatter dose from a typical monthly SPECT/CT workload on a new GE Infinia Hawkeye 4 system was measured by means of a dosimeter badge survey. Doses were found to be well within regulatory limits. Scatter dose rates attributable solely to the CT component were also measured experimentally for comparison with the manufacturer's isocontour plot. The measured CT dose rate isocontours were found to be of similar magnitude and location to those quoted by the manufacturer. Data generated from this study will form a valuable baseline for the monitoring of equipment performance and for future scatter radiation measurements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.