Abstract

Using a mesh of 30 thermoluminescent dosemeters, adults' patient skin doses were measured for 99 coronary angiography (CA) and 89 percutaneous coronary interventions (PCI) performed in three Lebanese hospitals. Average peak skin dose (Dskin,max) were 152 mGy (range: 16-1144) for CAs and 576 mGy (range: 7-3361) for PCIs. While only four patients had a Dskin,max value exceeding the 2 Gy threshold for skin injuries, several patients had skin dose values above 1 Gy at several distinct locations proving that Dskin,max alone is not sufficient for repetitive procedures; 2D dose maps are required instead. Dskin,max correlated well with total air kerma-area product (PKA,T) for PCI in Hospitals 1 and 2 (R = 0.91 and 0.76, respectively) enabling the setup of an alert level at PKA,T = 240 and 210 Gy cm2, respectively, corresponding to a Dskin,max of 2 Gy. This was not possible for Hospital 3 due to weak correlations between Dskin,max and PKA,T.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.