Abstract

The construction of a new vacuum system fabricated for pressure calibration in UHV/XHV region is explained together with the measurement of outgassing rate from an XHV chamber used in the system. The outgassing rate was measured using a test chamber by a throughput method after each treatment of the chamber surface such as electrolytic polishing, pre-baking in vacuum, and TiN coating. It was found that the outgassing rate of 3.7×10 -9 Pa m s -1 after electrolytical polishing could be reduced to 1.0×10 -13 Pa m s -1 after pre-baking twice followed by TiN coating. The outgassing rate was also measured by the pressure rise method using the whole XHV chamber, and the result was compared with that obtained with the test chamber.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.