Abstract

A method that provides for direct in vitro measurement at 37°C of ionized magnesium in biological fluids using the Abbot bichromatic analyzer, a fully automated instrument is described. The metallochromic dye Eriochrome blue SE was used to measure ionized magnesium levels in Mueller-Hinton agar fluids. Formulation of an appropriate blank was found necessary and was made by treating agar fluids with Chelex 100 to remove divalent cations. The influence of pH in the biological range was examined and poses a significant but manageable source of error. The method allowed an appropriate assessment of the role of ionized magnesium in the susceptibility testing of Pseudomonas aeruginosa on Mueller-Hinton agar with the aminoglycoside antibiotics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.