Abstract
SummaryA feasibility study was carried out to assess the potential of near infrared (NIR) reflectance spectroscopy for the measurement of fat and sucrose in dry cake mixes. The calibration of the NIR instrument was carried out in a research laboratory and then assessed under quality control conditions in the factory laboratory. It was possible to measure fat with an accuracy (±2σ) of ±3.4% for products with a fat content of 8–25% compared with +0.76% for the precision of the Soxhlet procedure. In the case of sucrose the accuracy was ±5.4% for products with a sucrose content of 10–40% compared with ±2.0% for a high pressure liquid chromatography (hplc) method. It must be concluded, therefore, that while NIR offers a quicker, simpler method of quality control, this is at the expense of accuracy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Food Science & Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.