Abstract

Purpose: Determine the exposure to personnel accrued while performing shielding evaluations of diagnostic facilities. Methods : Shielding inspections were performed at three sites: a small animal imaging facility, an imaging center consisting of general x-ray, CT, and PET/CT rooms, and a cardiac catheterization lab. The small animal facility was evaluated using two 20 mCi Tc-99m uncollimated sources. The imaging center was evaluated using a 20 mCi source of F-18 for the PET/CT room and a collimated 25 mCi source of Tc-99m for the CT and x-ray rooms. The catheterization lab was evaluated using an uncollimated 35 mCi vial of Tc-99m. Electronic personal dosimeters (Rados RAD-60R, RADOS Technology Oy Turku, Finland) were worn by each of the personnel involved in performing shielding inspections. The cumulative exposure reading from the dosimeter for each surveyor was recorded. Results: The surveyor who positioned the open source at the small animal facility had an exposure reading of 3.5 mR in 5.5 hours. Those who were outside the room surveying the barriers had readings of 0 and 0.1 mR in 5 hours. For the cath lab, the surveyor positioning the source had a reading of 0.4 mR in 30 minutes, while those outside had readings of 0 mR. At the imaging center all personnel surveying the PET/CT room with an open F-18 source had readings of 2.0, 1.1, and 2.9 mR in 4 hours. Surveys of the CT and general radiography rooms with a collimated Tc-99m source had readings of 1.2, 0.9, and 0.2mR in 5 hours. Conclusion : The surveyors who were in the room with the sources had the highest exposures. Using a higher energy source (F-18) also led to higher exposures. By using a collimated source, the time to measure each barrier for penetrations was increased, but the surveyors’ exposures were lower. --------------------------------------- Cite this article as: Page L, Dodge C, Li G, Wendt R. Measurement of exposure to personnel performing shielding evaluations. Int J Cancer Ther Oncol 2014; 2(2):020230. DOI: 10.14319/ijcto.0202.30

Highlights

  • Determine the exposure to personnel accrued while performing shielding evaluations of diagnostic facilities

  • Shielding inspections were performed at three sites: a small animal imaging facility, an imaging center consisting of general x-ray, CT, and PET/CT rooms, and a cardiac catheterization lab

  • The imaging center was evaluated using a 20 mCi source of F-18 for the PET/CT room and a collimated 25 mCi source of Tc-99m for the CT and x-ray rooms

Read more

Summary

Introduction

The surveyor who positioned the open source at the small animal facility had an exposure reading of 3.5 mR in 5.5 hours. Those who were outside the room surveying the barriers had readings of 0 and 0.1 mR in 5 hours. The surveyor positioning the source had a reading of 0.4 mR in 30 minutes, while those outside had readings of 0 mR. At the imaging center all personnel surveying the PET/CT room with an open F-18 source had readings of 2.0, 1.1, and 2.9 mR in 4 hours. Surveys of the CT and general radiography rooms with a collimated Tc-99m source had readings of 1.2, 0.9, and 0.2mR in 5 hours

Methods
Results
Conclusion