Abstract

We characterized the energy response of personal dose equivalent (Hp(10) in mrem) and the contribution of backscatter to the readings of two electronic personal dosimeter (EPD) models with radionuclides commonly used in a nuclear medicine clinic. The EPD models characterized were the RADOS RAD‐60R, and the SAIC PD‐10i. The experimental setup and calculation of EPD energy response was based on ANSI/HPS N13.11‐2009. Fifteen RAD‐60R and 2 PD‐10i units were irradiated using 99mTc, 131I, and 131I radionuclides with emission energies at 140 keV, 364 keV, and 511 keV, respectively. At each energy, the EPDs output in Hp(10) [mrem] were recorded with 15 inch thick PMMA to simulate backscatter form the torso. Simultaneous free‐in‐air exposure rate measurements were also performed using two Victoreen ionization survey meters to calculate the expected EPD Hp(10) values per ANSI/HPS N13.11‐2009. The energy response was calculated by taking the ratio of the EPD Hp(10) readings with the expected Hp(10) readings and a two‐tailed z‐test was used to determine the significance of the ratio deviating away from unity. The contribution from backscatter was calculated by taking the ratio of the EPD Hp(10) readings with and without backscatter material. A paired, two‐tailed t‐test was used to determine the significance of change in EPD Hp(10) readings. The RAD‐60R mean energy response at 140 keV was 0.85, and agreed to within 5% and 11% at 364 and 511 keV, respectively. The PD‐10i mean energy response at 140 keV was 1.20, and agreed to within 5% at 364 and 511 keV, respectively. On average, in the presence of acrylic, RAD‐60R values increased by 32%, 12%, and 14%, at 140, 364, and 511 keV, respectively; all increases were statistically significant. The PD‐10i increased by 25%, 19%, and 10% at 140 keV, 364 keV, and 511 keV, respectively; however, only the 140 keV measurement was statistically significant. Although both EPD models performed within the manufacturers' specifications of ±25% in the energy ranges used, they fell outside of our criteria of 10% at lower energies, suggesting the need to calculate energy‐dependent correction factors, depending on the intended EPD use.PACS numbers: 87.53.Bn, 87.55.N‐, 87.57.U‐

Highlights

  • 424 Meier et al.: Characterization of electronic personal dosimeters are often used by pregnant technologists in nuclear medicine departments to monitor radiation exposure during pregnancy

  • Hp(10) energy response At all energies for both electronic personal dosimeter (EPD) models, the energy response of the EPDs for both models did not diverge from unity by greater than 25%, and all were operating within the manufacturers’ energy response specifications of ± 25% for energies between 60 keV and 3 MeV

  • Hp(10) energy dependency Characterizations of EPDs have been conducted by others, and they have shown that there is a very broad performance in the energy response of the many EPDs that are available on the market with respect to radiation type, photon energy, photon beam quality, and the dose rate.(3-5) all evaluated EPDs operated within the manufacturers’ energy response specifications of ± 25%, our results show that both of the EPD models investigated exhibited an energydependent response, at least between 140 to 511 keV

Read more

Summary

Introduction

424 Meier et al.: Characterization of electronic personal dosimeters are often used by pregnant technologists in nuclear medicine departments to monitor radiation exposure during pregnancy. The real-time (e.g., daily) dose measurements provide information that can be used to proactively modify work responsibilities and schedules of the pregnant technologist. This flexibility provides a distinct advantage to relying on monthly readings of whole body dosimeters. EPDs can be used to obtain a better understanding of transient dose received during routine testing of clinical nuclear medicine equipment. Even though EPDs are not used as the dose of record but as a dose estimate in these situations, a miscalibrated or faulty EPD could lead to the premature or tardy change of work responsibilities for a pregnant technologist

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call