Abstract

Using a luminescence spectrometer as a platform, a system of fibre-optic probes was created that allows full colour characterisation, fluorescence and phosphorescence spectra to be recorded in diffuse reflectance and in transmission from thick or thin film arrays of combinatorial samples of diameter down to 2 mm and from liquids. An integrating sphere is not required and the method is more versatile than the instrument's fibre-optic plate reader which has conjoined fibre bundles set at a fixed angle. Incident and detected light is routed via separate optical fibre bundles which remain stationary above or below a two-axis table. The validation and calibration are described. A library of 25 members was scanned for both diffuse reflectance (colour) and fluorescence in less than an hour. The method thus combines techniques that conventionally rely on different instruments and makes them amenable for high throughput libraries.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.