Abstract
For many liquid materials and biological media, their complex permittivity varies with frequency in the P-band (100-1000MHz). In this paper, a re-entrant coaxial cavity resonance test system with intensive multi-frequency points testing capability is designed for the P-band to realize the test of the complex permittivity of P-band materials. For this system, a test algorithm has been written, and on the basis of this system, a perturbation block has been designed, using stepper motors to control the up and down movement of the perturbation block to achieve frequency tunability of the cavity, enabling more P-band frequency test data to be obtained. The experimental results show that the system is suitable for testing the electrical parameters of P-band biological media and liquid materials and for testing solid materials in aerospace, electromagnetic shielding, and other fields where there is a great demand for P-band testing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.