Abstract

We have developed, as a national standard, a windowless multiwire proportional counting (MWPC) system for gross measurements of α - and β -emitting radionuclides at an uncertainty level below 1%. Various detector characteristics have been studied and several working parameters of the system have been determined in order to achieve an optimized detection capability. The detector is commissioned through an international comparison of the β emission rate from an extended area 36 Cl source. Correction for the dead time loss is made by the non-extendable dead time method. A linear extrapolation of pulse height spectrum is employed to correct for the loss in the detection efficiency due to discriminating against the events formed from a small number of original ion pairs. The surface emission rate was measured to be 1.84 × 10 3 counts/s with the relative combined standard uncertainty of 0.46%. The measured uncertainty indicates that the constructed MWPC is capable of serving as a national standard.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.