Abstract
This study investigates potential non-radiation hazards, specifically Ozone (O3) production, during Linear Accelerator (LINAC) electron mode radiotherapy. The research uses experimental measurements to determine Ozone concentration in the LINAC patient waiting room and control room. Measurements are taken assuming a ±2-hour delay in one working day, using 9 MeV and 12 MeV energy, 400 MU dose rate, and illumination angles of 0o, 90o, and 270o. Maximum Ozone concentrations in the LINAC patient waiting room and control room are found to be 6.6 ppb (12 MeV) and 8.3 ppb (12 MeV), respectively. These concentrations fall below the chemical threshold limit and are deemed safe for human exposure. Notably, potentially detectable Ozone levels are observed in the LINAC banker. Overall, this research highlights the importance of monitoring Ozone levels to ensure the safety of both patients and personnel in LINAC facilities.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.