Abstract

1H-MRS, which is a method of measuring fat and iron deposits in the liver, has a single voxel spectroscopy imaging method called high-speed T2-corrected multiecho (HISTO) based on the STEAM method. If HISTO measurement accuracy is high, it will be one of the biomarkers, so the measurement accuracy was examined. We imaged a self-made phantom simulating a known amount of fat and iron deposition in the liver. A regression analysis of HISTO-measured values for hepatic fat phantom was performed. In addition, regression analysis was performed on the hepatic iron phantom for the R2 values obtained from the T2 values measurement and the actually measured value of the HISTO. From the regression analysis, the correlation coefficients were 0.999, 0.992 and 0.977, respectively. The measurement accuracy of HISTO is high, and the reliability of the obtained fat and iron deposit value is high.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.