Abstract

The performance of HPLC pumps is determined by applying Hagen-Poiseuille's law. As restrictor a fused silica column is used. The pulsation height and frequency is measured via a pressure transducer. Fourier transformation of pulsation data permits the characterization of pump identity. The absolute calibration of HPLC detectors and the determination of their linear and dynamic range as well as the noise level can be achieved by a single measurement applying the exponential dilution flask. Wavelength accuracy of variable wavelength detectors can be determined by comparing spectral data of aromatic hydrocarbons from the literature with those determined with the detectors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.