A MCNP model of a252Cf source-based prompt gamma-ray neutron activation analysis (PGNAA) system for bulk sample analysis was benchmarked. The indium foils activation and aqueous samples containing chlorine were measured using the PGNAA system, and the experimental data compared to the simulations. The discrepancies between the measurements and simulations are within 6% and 20% for thermal neutron flux and prompt gamma-ray response, respectively. The results show that the simulations agree well with the measurements, and demonstrate the proposed MCNP model can represent the system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.