Abstract

In this work, a method was proposed for efficiency calibration of HPGe detector employed in a prompt gamma-ray neutron activation analysis (PGNAA) system for large aqueous sample analysis. Experimental measurements of a reference surface gamma-ray source were combined with efficiency transfer method to obtain efficiency of a volumetric source. Then efficiency curve over the energy range 0.5–8.5 MeV were determined with prompt gamma-rays of chlorine. To evaluate the performance of the calibration curve, an internal standard method was used to determine the aqueous sample containing manganese based on the calibration curve. The 7.058 MeV and 7.243 MeV manganese gamma-ray peaks were used to calculate the mass of manganese. The relative deviations of calculated values and actual value were 5.0% and 6.5%, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.