Abstract

We have quantitatively evaluated the usable thickness of specimens in scanning transmission electron microscopy (STEM) at 1 MV using a wedge-shaped Si(110) single crystal including artificially introduced high-density dislocations. The width of dislocation images was employed as a criterion for the quantitative evaluation of usable thickness. Superior usable thickness in STEM than in TEM was found; the obtained results were 14.7 µm for STEM and 5.8 µm for TEM. In particular, in STEM, dislocations can be observed as thin lines with 10–15 nm width in the thickness range up to 10 µm. The latest high-voltage STEM is useful for imaging crystal defects in thick semiconductors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.