Abstract

Inverse Monte Carlo (IMOC) reconstruction for SPECT uses a Maximum Likelihood (EM) estimator with detection probabilities generated as Monte Carlo solutions to the photon transport equation (PTE). To evaluate the behavior of the iterative EM algorithm, reconstructions from experimental projection data for up to 1000 iterations were examined. Compensation for scatter and attenuation was achieved by including those effects in the PTE. For uniform activity distributions, noise increased monotonically with iteration. With line sources included, both noise and resolution improved between 1 and 30 iterations after which resolution was slightly improved at the expense of noise. Contrast (for a nonactive region surrounded by activity) improved from 0.32 at 5 iterations to 0.90 at 200 iterations, and 0.97 at 1000. Uncertainty in the measurement increased due to increased noise in the active region.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.