Abstract

The quality is essential for radiation use in medicine and the beam quality index is the basic parameter to periodically check the normal functioning of Linac head in radiotherapy treatment. It is recommended by IAEA protocols TRS-398 based on absorbed dose in water. The PDD method is used as a basis of the parameterization of the photon beam quality for predicting its variation with beam energy and field size. The objective of this work is to establish a mathematical law for predicting and checking the beam dosimetry quality index according to field size and beam energy based on IAEA TRS-398 protocol.For an easier and more reliable procedure determination of the beam quality based on TRS 398, two empirical laws were therefore established with an accuracy better than 2%. They can serve the basic floor to medical physicist to verify and to control the dosimetry output quality at arbitrary field size. Our findings aim to facilitate the dosimetry quality control that set in use according to current conditions for checking out the radiotherapy efficiency and the safety inside the treatment room.