Abstract
In the present study, twelve explanted mechanical heart valves (MHVs)with pyrolitic carbon tilting disc and 14 bileaflet MHVs were analyzed to investigate the effects of material properties on valve performance and patients' general health conditions. Optical and scanning electron microscopy was used to investigate material imperfections, wear patterns or damages to housing and occluder components. All analyzed tilting disc valves exhibited wear effects, particularly due to abrasion and impact to both disc and housing. Wear of pyrolitic carbon disc and housing did not influence their in vivo performance. In the bileaflet MHVs, breakaway of the pyrolitic carbon coating sometimes caused malfunctioning and required surgical retrieval of the valve. In all cases, occurrence of clinical symptoms was more likely when wear effects were located in critical areas. The study supports a correlation between the properties of the MHVs material and patients' symptoms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.