Abstract

The recent upgrade to the MAST YAG Thomson scattering while enhancing the diagnostic capabilities increased the complexity of the system. There are eight YAG lasers now operational, doubling the number from the previous setup. This means alignment between each laser individually and reference points is essential to guarantee data quality and diagnostic reliability. To address this issue an alignment system was recently installed. It mimics the beams alignment in MAST by sampling 1% of the laser beam that is sent into a telescope which demagnifies by a factor of 8. The demagnified beam is viewed with a CCD camera. By scanning the camera the profile and position of the beams in the scattering zone and in a range of several meters inside MAST can be determined. Therefore alignment is checked along the beam path without having to sample it inside the vessel. The experimental apparatus and test procedures are described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.