Abstract

When coupling capillary electrophoresis with postcolumn detection methods, such as mass spectrometry, the presence of postcolumn band broadening must be considered. The band broadening effects introduced by junction-at-the-tip CE-MS interfaces using a postcolumn micro flow-through vial are investigated by studying the hydrodynamic flow patterns and mass transport process inside the micro vial at the end of the CE separation capillary. Simulation results obtained by solving the Navier-Stokes and mass balance equations provide insights into the velocity field and concentration distribution of the analytes in the micro vial and demonstrate that, with a low flow rate of chemical modifier solution, the laminar flow streams confine the analyte molecules to the central part of the micro vial and thus maintain major features of the peak shapes. Peaks detected by UV and MS under similar experimental conditions were compared to verify the numerical prediction that the main features of the UV peak can be retained in the MS peak. Experiments also show that band broadening can be minimized when an appropriate chemical modifier flow rate is selected.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.