Abstract

A new strategic masking technique, namely the ABC method, has been developed. In performing this method of measuring thresholds of bone-conduction, the vibrator is placed at the forehead with both ears occluded by air-conduction earphones. One of the earphones is for masking noise and the other is a dummy which balances out the occlusive effect of the test ear against the nontest ear. The ABC method is based on the ABC rule that, in bone-conduction testing, the effective masking noise level necessary to block out the nontest ear can be calculated by a simple equation: right AC (A) + left AC (B)--unmasked BCu (C) under the assumption that the BCu belongs to the nontest ear. In some cases of hearing loss, the above noise level might produce overmasking, then an additive safety noise level, BCu + Interaural Attenuation, is employed. This method offers testers step by step directions which consist of indications of the noise level and a criterion for determining whether the measured bone-conduction is free from cross hearing and overmasking for the given configuration of air-conduction of both ears, BCu, and the masking noise level. Compared to the well known Plato method, in which measurements of thresholds are repeated at several masking noise levels in order to find a single bone-conduction threshold, the ABC method can essentially find the threshold at only one masking noise level. Therefore the ABC method makes it possible to save a great deal of time in performing bone conduction testing.