Abstract

This study evaluated the accuracy of marginal and internal fit between the all-ceramic crowns manufactured by a conventional double-layer computer-aided design/computer-aided manufacturing (CAD/CAM) system and a single-layer system. Ten standardized crowns were fabricated from each of these two systems: conventional double-layer CAD/CAM system (Procera) and a single-layer system (Cerec 3D). The copings and completed crowns were seated on the abutments by a special device that facilitated uniform loading, and the marginal discrepancies were measured. Internal gaps were also measured using a low-viscosity silicone material. Marginal discrepancies of Procera copings were significantly smaller than those of Procera crowns and Cerec 3D crowns (p < 0.05), but Procera crowns and Cerec 3D crowns did not differ significantly from each other (p > 0.05). On internal gaps, Cerec 3D crowns showed significantly larger internal gaps than Procera copings and crowns (p < 0.05). Within the limitations of this study, the single-layer system demonstrated acceptable marginal and internal fit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.