Abstract

This study aims to assess the dimensional accuracy of complete denture bases fabricated from different CAD/CAM technologies and a conventional method, including milling (CNC), PolyJet (PJ), laser sintering (SLS), digital light processing (DLP), and injection molding (IM). It also examines the influence of the removal of technology-specific connectors or support structures when present. Denture base surfaces were digitized using a laboratory scanner, and virtual measurement points were calculated with tetrahedral reference geometries. Defined distances were measured in all spatial directions and compared to design data (p<0.05), revealing significant differences in sagittal (p=0.004), transversal (p<0.001), and vertical (p<0.001) dimensions. Connector removal had no significant impact for CNC but significantly affected DLP. All technologies yielded clinically acceptable results, with CNC milling demonstrating the best overall outcome.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.