Abstract
We present an optical coherence tomography (OCT) based method for mapping transverse red blood cell (RBC) flow speed at capillary. This OCT velocimetry utilizes a quantitative laser speckle temporal contrast analysis that estimates reliable speckle decorrelation time from the observed speckle contrast, which is related to microcirculatory flow velocity. For capillary speed measurement, we employ a home-built 1.3 µm MHz swept-source OCT (SS-OCT) system that can acquire OCT B-frames at a rate of 1.7 kHz. From the multiple B-frames obtained at the same location, intensity profiles with time-varying OCT speckle contrast are extracted at single capillaries using a capillary binary mask and then the transverse flow speed is calculated by adapting the profiles to the speckle contrast analytic model. Finally, a 3D speed map can be achieved for OCT volume imaging. To validate this method, we perform a systematic study using both phantom and in vivo rodent models. Result shows that our method is effective to measure transverse capillary flow speed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.