Abstract

This article demonstrates a method to salvage an implant that has been damaged or is no longer usable because of a fractured screw that cannot be removed. In the first situation the patient had a Locator abutment that fractured during torqueing of the abutment. Despite the attempt to remove the fractured screw, it was not retrievable, and the internal threads of the implant were damaged in the attempt to remove the screw. Rather than removing the implant or covering it and sleeping the implant, a custom cast post was used to restore the implant to function. A second example describes a patient with a fixed implant crown. The implant was damaged during attempts to remove the fractured screw. A custom post was fabricated and cemented, and a new crown was fabricated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.