Abstract
Malfunctions of the dose monitoring systems of two linear accelerators were studied. A variation of plus or minus 2.5% over an air density correction factor range from 1.02 to 1.07 was measured, suggesting that the ionization-type monitor chambers were open to the atmosphere. An average variation of minus 2.0% was present between early morning calibrations and those performed later in the day. Any variation caused by changes in control console temperature was within experimental error. This type of malfunction can lead to errors in delivered doses of 5%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.