Abstract
A novel dual wavelength procedure is described for the elimination of baseline pump noise in spectrophotometric post-column reaction detection systems in high-performance liquid chromatography. The method is applicable if there is some overlap between the spectrum of the analyte reaction product and the spectrum of the excess reagent. One wavelength is chosen to monitor the analyte reaction product and the other wavelength is set close to the isosbestic point. After simple arithmetic manipulation of the data using commonly available computer software, the pump noise is subtracted from the analyte baseline, allowing much higher sensitivity settings to be used. The two examples shown are taken from the ion chromatography of trace metals using colorimetric post-column reaction detectors, where close to a 20-fold increase in signal to noise ratio was obtained when using the dual wavelength procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.