Abstract

Closed-loop insulin delivery systems, which integrate continuous glucose monitoring (CGM) and algorithms that continuously guide insulin dosing, have been shown to improve glycaemic control. The ability to predict future glucose values can further optimize such devices. In this study, we used machine learning to train models in predicting future glucose levels based on prior CGM and accelerometry data. We used data from The Maastricht Study, an observational population-based cohort that comprises individuals with normal glucose metabolism, prediabetes, or type 2 diabetes. We included individuals who underwent >48h of CGM (n = 851), most of whom (n = 540) simultaneously wore an accelerometer to assess physical activity. A random subset of individuals was used to train models in predicting glucose levels at 15- and 60-minute intervals based on either CGM data or both CGM and accelerometer data. In the remaining individuals, model performance was evaluated with root-mean-square error (RMSE), Spearman's correlation coefficient (rho) and surveillance error grid. For a proof-of-concept translation, CGM-based prediction models were optimized and validated with the use of data from individuals with type 1 diabetes (OhioT1DM Dataset, n = 6). Models trained with CGM data were able to accurately predict glucose values at 15 (RMSE: 0.19mmol/L; rho: 0.96) and 60 minutes (RMSE: 0.59mmol/L, rho: 0.72). Model performance was comparable in individuals with type 2 diabetes. Incorporation of accelerometer data only slightly improved prediction. The error grid results indicated that model predictions were clinically safe (15 min: >99%, 60 min >98%). Our prediction models translated well to individuals with type 1 diabetes, which is reflected by high accuracy (RMSEs for 15 and 60 minutes of 0.43 and 1.73 mmol/L, respectively) and clinical safety (15 min: >99%, 60 min: >91%). Machine learning-based models are able to accurately and safely predict glucose values at 15- and 60-minute intervals based on CGM data only. Future research should further optimize the models for implementation in closed-loop insulin delivery systems.

Highlights

  • The increasing prevalence of diabetes entails an increase in debilitating complications, such as retinopathy, neuropathy, and cardiovascular disease [1,2,3]

  • Models trained with continuous glucose monitoring (CGM) data were able to accurately predict glucose values at 15 (RMSE: 0.19mmol/L; rho: 0.96) and 60 minutes (RMSE: 0.59mmol/L, rho: 0.72)

  • Our prediction models translated well to individuals with type 1 diabetes, which is reflected by high accuracy (RMSEs for 15 and 60 minutes of 0.43 and 1.73 mmol/L, respectively) and clinical safety (15 min: >99%, 60 min: >91%)

Read more

Summary

Introduction

The increasing prevalence of diabetes entails an increase in debilitating complications, such as retinopathy, neuropathy, and cardiovascular disease [1,2,3]. One of the most promising developments to minimize hyperglycaemia and hypoglycaemia–and, to increase time in range–in individuals with diabetes who require insulin treatment is a closed-loop insulin delivery system ( known as the artificial pancreas). Such a system integrates continuous glucose monitoring (CGM), insulin (with or without glucagon) infusion, and a control algorithm to continuously regulate blood glucose levels [6, 7]. Closed-loop insulin delivery systems, which integrate continuous glucose monitoring (CGM) and algorithms that continuously guide insulin dosing, have been shown to improve glycaemic control. We used machine learning to train models in predicting future glucose levels based on prior CGM and accelerometry data

Objectives
Methods
Results
Conclusion