Abstract

Lymphocyte 51Cr release and [ 3H]thymidine uptake assays were evaluated with respect to measurement of sample radioactivity using the flat-bed scintillation counter. 51Cr lysates were spotted onto a glass fibre filter sheet while [ 3H]thymidine-labelled cells were filtered onto a similar sheet using a cell harvester. The 96 samples were rapidly processed for counting, without removal of individual sample areas. Either form of preparation showed good linearity of count rate with the quantity of material on the filter. Reproducibility was good; the coefficient of variation for 96 samples being within 5%. The low background and high efficiency of this counter results in increased assay sensitivity and allows considerable economies in materials to be made. A commercial version of the counter has six counting heads permitting a high rate of sample throughput.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.