Abstract
Patients undergoing mechanical lung ventilation are at risk of lung injury. A noninvasive bedside lung monitor may benefit these patients. The Inspired Sinewave Test (IST) can measure cardio-pulmonary parameters noninvasively. We propose a lung simulation to improve the measurement of pulmonary blood flow using IST. The new method was applied to 12 pigs' data before lung injury (control) and after lung injury (ARDS model). Results using the lung simulation shown improvements in correlation in both simulated data (R2 increased from 0.98 to 1) and pigs' data (R2 increased from <0.001 to 0.26). Paired blood flow measurements were performed by both the IST (noninvasive) and thermodilution (invasive). In the control group, the bias of the two methods was negligible (0.02L/min), and the limit of agreement was from -1.20 to 1.18 L/min. The bias was -0.68 L/min in the ARDS group and with a broader limit of agreement (-2.49 to 1.13 L/min).Clinical Relevance- the inspired sinewave test can be used to measure cardiac output noninvasively in mechanically ventilated subjects with and without acute respiratory distress syndrome.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.