Abstract

Soft X-ray response of X-ray Imaging Spectrometers (XIS) for the Astro-E satellite is measured with a grating spectrometer system at Osaka. First, relation between incident X-ray energy and output pulse height peak (E-PH relation) is examined with an SX grating. It is found that jump in the E-PH relation around Si-K edge is at most 2.7 eV. Second, quantum efficiency (QE) of the XIS in 0.4 - 2.2 keV range is measured relatively to the reference CCD of which absolute QE was calibrated with a gas proportional counter. The QE is fitted with a model in which CCD gate structures are considered. Systematic error on the QE results is estimated by referring an independent measurement. Third, tuning and improvement of the response function is performed. We employ six components to reproduce the response profile of the XIS. In this paper, improvement of one component which is originated in the events absorbed in the channel-stop is presented. Nevertheless, Astro-E was lost due to the launch failure. We overview the XIS project in its flight model phase, modified points of the design, problems and solutions etc., in order to be utilized in a possible recovery of the satellite.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.