Abstract

A new sensor for respiratory rate monitoring based on optical fiber technique was simultaneously compared with an acoustic sensor and a transthoracic impedance plethysmograph during normoventilation in the respiratory rate range of 9 - 17 breaths per minute. The response characteristics of the optical sensor were then measured during simulation of central apnoea and tachypnoea. Visual observation was chosen as the reference method for monitoring the respiratory rate. The measurements were performed in ten healthy volunteers and the respiratory signals recorded on an analogue tape and strip-chart recorder and analyzed off-line. The response characteristics of the fiber optic sensor corresponded well with those of the acoustic sensor and impedance plethysmograph. All three methods responded rapidly to a simulated apnoeic event.© (1995) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.