Abstract

Visual observation was chosen as the reference method for measuring the respiratory rate in ten healthy volunteers. The new fibre-optic and acoustic sensors were simultaneously compared with capnography and transthoracic impedance plethysmography during normoventilation in the respiratory rate range of 6-24 breaths per minute and at a fixed respiratory rate of 13 breaths per minute. In addition a simulation of central apnoea was performed. All the measurements were recorded on an analogue tape recorder and a strip-chart recorder and analysed off line. The analyses of the recordings were performed by a person who was unable to see the monitoring systems. There was no discrepancy in the results of these methods. Each of the methods responded rapidly to an apnoeic event. The new fibre-optic and acoustic sensors correlate well with more traditional methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.