Abstract

Charge transfer efficiency (CTE) test methods are reviewed, and the results and conclusions of the tests are given. The test methods have been utilized to describe the CTE characteristics of the Tektronix 1024 by 1024 CCD to optimize low dark current, low readout noise, and high CTE at low signal levels. CTE modelling is described, and three test methods are set forth and compared. The Fe-55 X-ray response method utilizes the response of a CCD to X-ray photons from the radioactive source Fe-55. The extended pixel edge response method employs the measurement of the charge lost to successive pixels by a known initial signal as it is shifted through the array. The charge injection method consists of charge injection through the output amplifier reset transistor. These measurements were performed on several devices with known CTEs. The CTEs are found to be in agreement for the three methods, making application and test requirements the principal criteria for their use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.