Abstract

The detection limit of conducting polymer (CP) poly(pyrrole)-based potentiometric chloride-sensitive electrodes was lowered over 3 orders of magnitude, applying anodic current of density in the range microamperes per centimeter squared. This effect was attributed to compensation of doping chloride ion leakage from the membrane into adjacent solution layer, caused by self-discharge of CP. This method was successfully applied to lower the detection limit of all-solid-state chloride-selective potentiometric sensors containing plastic, poly(vinyl chloride) chloride ion-selective membrane, and conducting poly(pyrrole) film applied as ion-to-electron transducer. The control of the leakage of chloride ions from the CP transducer phase to the plastic membrane resulted in linear response of ion-selective electrode shifted down to lower activities. A 2 orders of magnitude lowered detection limit, equal to 4 × 10-7 M, was achieved for current densities corresponding to extended linear range of CP film electrode, i.e...