Abstract

We describe a low-pressure plug-flow microreactor system for investigating reaction kinetics on high surface area catalysts over a wide pressure range, covering the so-called pressure gap. Well defined gas flows with rates spanning over six orders of magnitude, from 100Nml∕min down to 0.5Nμl∕min, are produced using a capillary system for gas flow control. This allows us to measure reaction rates at constant contact time, despite widely differing pressures. The performance of the reactor system is demonstrated by test measurements on the kinetics for CO oxidation over a Au∕TiO2 catalyst, covering a pressure range from 1000mbarto10−1mbar total pressure, equivalent to CO partial pressures of 10mbarto10−3mbar.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.