Abstract

Describes two custom integrated circuits which were developed for an implantable pulsed Doppler ultrasonic blood flowmeter. Prime design goals were a minimum circuit volume, minimum power consumption, and operation at low supply voltages. The first of the two IC's performs system timing functions and produces the ultrasonic transmit burst. It can deliver up to 40 mW of peak output power at 50 percent efficiency and requires 3.7 mW standby power. The second IC, containing an RF amplifier, mixer and output amplifier, provides 54 dB conversion voltage gain for an 0.8 MHz bandwidth centered at 5.8 MHz, <3 dB noise figure, a dynamic range of 40 dB, and 1 /spl mu/s recovery time from a 1 V overload. This chip requires 2.7 mW power input.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.