Abstract

Abstract Ground conductivity meters, comprising a variety of coil–coil configurations, are intended to operate within the limits provided by a low induction number (LIN), electromagnetic condition. They are now routinely used across a wide range of application areas and the measured apparent conductivity data may be spatially assembled and examined/correlated alongside information obtained from many other earth science, environmental, soil and land use assessments. The theoretical behaviour of the common systems is examined in relation to both the prevailing level of subsurface conductivity and the instrument elevation. It is demonstrated that, given the inherent high level of accuracy of modern instruments, the prevailing LIN condition may require operation in environments restricted to very low (

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.