Abstract

IntroductionAccurate cardiac left ventricle (LV) delineation is essential to CT-derived left ventricular ejection fraction (LVEF). To evaluate dose-reduction potential, an anatomically accurate heart phantom, with realistic X-ray attenuation is required. We demonstrated and tested a custom-made phantom using 3D-printing, and examined the influence of image noise on automatically measured LV volumes MethodsA single coronary CT angiography (CCTA) dataset was segmented and converted to Standard Tessellation Language (STL) mesh, using open-source software. A 3D-printed model, with hollow left heart chambers, was printed and cavities filled with gelatinized contrast media. This was CT-scanned in an anthropomorphic chest phantom, at different exposure conditions. LV and “myocardium” noise and attenuation was measured. LV volume was automatically measured using two different methods. We calculated Spearmans’ correlation of LV volume with noise and contrast-noise ratio respectively om 486 scans of the phantom. Source images were compared to one phantom series with similar parameters. This was done using Dice coefficient on LV short-axis segmentations. ResultsPhantom “Myocardium” and LV attenuation was comparable to measurements on source images. Automatic volume measurement succeeded, with mean volume deviation to patient images less than 2 ml. There was a moderate correlation of volume with CNR, and strong correlation of volume with image noise. With papillary muscles included in LV volume, the correlation was positive, but negative when excluded. Variation of volumes was lowest at 90–100 kVp for both methods in the 486 repeat scans. The Dice coefficient was 0.87, indicating high overlap between the single phantom series and source scan. Cost of 3D-printer and materials was 400 and 30 Euro respectively. ConclusionBoth anatomically and radiologically the phantom mimicked the source scans closely. LV volumetry was reliably performed with automatic algorithms. Implications for practicePatient-specific cardiac phantoms may be produced at minimal cost and can potentially be used for other anatomies and pathologies. This enables radiographic phantom studies without need for dedicated 3D-labs or expensive commercial phantoms.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call