Abstract

The electrode-tissues interface (ETI) is one of the key issues for the safety, reliability and efficiency of implantable devices such as stimulators and sensors. The aim of this paper is to report an implantable telemetry device, based on a full custom integrated circuit (IC) to monitor the ETI. The proposed system performs various types of measurements, such as impedance spectroscopy, cyclic voltammetry, and galvanostatic double pulse method. Hence, the evolution of various electrochemical parameters of the ETI such as complex impedance, faradic resistance, double layer capacity, rheobase current, and chronaxy time, could be monitored long time after implantation. The full custom IC has been designed and fabricated with the CMOS 0.18μm technology. The circuit occupies a silicon area of 2mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> , and consumes less than 3mW. Characterisation and in-vitro experimental results validate the full functionalities of the implantable monitoring system including the custom IC.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.