Abstract

This paper reports long-term evaluation of a micropackage technology for an implantable MEMS pressure sensor. The all-polymer micropackage survived 160 days when subjected to accelerated lifetime testing at 85 °C in a 1% wt. saline solution. The package shows minimum effect on sensors' sensitivity and nonlinearity, which deviated by less than 5% and 0.3%, respectively. A 6-month in vivo evaluation of 16 MEMS-based pressure sensors demonstrated that the proposed micropackage has good biocompatibility and can protect the MEMS pressure sensor. To the best of our knowledge, these results establish new lifetime records for devices packaged using an all-polymer micropackaging approach.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.