Abstract

An experimental investigation to improve the lifetime of a discharge-excited ArF-excimer laser is presented. The three dominant factors restricting its lifetime are CF4 generation in the laser gas, color-center formation in the optics and input power density reduction due to electrode ablation. Copper electrodes were superior to nickel electrodes in regard to electrode ablation. A gas lifetime of more than 109 shots (about one month at 400 Hz) is shown for an ArF-excimer laser with a liquid-nitrogen trap and high-temperature zirconium alloy trap.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.