Abstract
Due to the radiosensitivity of paediatric patients to X-ray, it is necessary to survey the paediatric DRLs using size-specific dose estimates (SSDE). In the present study, we determined the local diagnostic reference levels (DRLs) for paediatric chest, head and abdomen-pelvis CT examinations and their Surview scans in Kermanshah city, Iran. For ≤1 year, 1-5 years, 5-10 years and 10-15 years the DRLs (mGy) based on SSDE were determined N/A, 6.00, 6.25, 8.27 for abdomen-pelvis, and 8.74, 7.45, 11.15, 10.45 for chest and 19.05, 18.33, 18.22, 20.14 for head examinations, respectively. The differences between body size and default phantom defined in CT scanners are significant and should be considered when determining the DRLs. Based on our findings, use of CTDIv and SSDE parameters for determining DRLs leads to significant different results in children; thus SSDE is suggested as a more accurate index than CTDIV for establishing DRLs in paediatric CT examinations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.