Abstract
PurposeThe purpose of this study was to determine local DRLs for children and adults undergoing intraoral dental examinations at the intraoral radiology units of the public hospitals in Cyprus. MethodsMeasurements were made on all the twenty intraoral X-ray units of the public hospitals in Cyprus with the intention to establish the local DRLs for all the possible intraoral X-ray examinations for children and adults. All units are film based. The measurements were made by a Dose Area Product (DAP) meter (GAMMEX RMI 841-RD) placed at the surface of the dental unit’s X-ray shaping cone (FSD 20cm). A diagnostic radiology dosimeter (Dosimax Plus A) was also placed at an FSD of 100cm to compare the dose reading between the two dosimeters. ResultsDRLs were established at the 3rd quartile for 7 exposure settings corresponding to 12 types of teeth (Adult and children mandibular and maxillary incisor, premolar and molar) with values of 197, 163, 128, 102, 81, 65 and 49mGycm−2 and 7.23, 5.94, 4.75, 3.68, 3.10, 2.41 and 1.88mGy for benchmark nominal exposure times of 1000, 800, 640, 500, 400, 320 and 250ms respectively, at a nominal exposure voltage of 70kVp. ConclusionsThe local DRLs of the present study compare well with other similar published DRLs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.