Abstract

This study was performed to evaluate the efficacy of a novel computed tomography (CT) liver detection algorithm (LDA), which allows for targeted increase of radiation dose to the upper abdomen, on image quality of the liver. We retrospectively evaluated the LDA by comparing 40 consecutive patients who had portal venous CT abdomen performed without use of the algorithm, to 40 patients in whom the algorithm was used. Image quality was assessed objectively by comparing the standard deviation (SD) of attenuation values in Hounsfield units (HU) of the abdominal organs. Qualitative analysis was performed by two blinded radiologists who independently graded the image quality of abdominal organs There was significant noise reduction in the liver (P<0.001) and spleen (P<0.001) in the LDA group compared to the conventional group. There was also a significant improvement in image quality of the liver (P<0.001), kidney (P<0.001), spleen (P<0.001), pancreas (P<0.001), and psoas (P=0.005) in the LDA group compared to the conventional group. Overall dose between the two groups was similar. This liver detection algorithm improves the subjective image quality of upper abdominal organs, in particular the liver, without increasing overall radiation dose.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.