Spuriously high results using the Abbott Architect enzymatic creatinine assay were noted to be particularly associated with very small sample volumes. This led us to query the effect of under-filling lithium heparin tubes on the measured enzymatic creatinine result. Blood was provided by 5 laboratory personnel and then decanted into 5 x1.2mL Sarstedt S-Monovette tubes, giving final blood volumes of 200, 400, 600, 800 and 1200μL. Plasma was analysed using Abbott Architect Jaffe, enzymatic creatinine, Beckman Coulter (AU500) enzymatic creatinine and Roche (Cobas c702) enzymatic creatinine assays. Saline was also added to Sarstedt 1.2mL and Teklab 2mL tubes and analysed using the Abbott Jaffe and enzymatic creatinine methods. Increasing degrees of under-fill were associated with greater over-estimation of creatinine using the Abbott enzymatic assay, but no difference was noted using Jaffe methodology on the same platform or enzymatic assays provided by Roche or Beckman. On average, creatinine was 40.6% (+27.7μmol/L) higher when only 200μL of blood was present in the tube. Small volumes of saline added to lithium heparin tubes measured significant creatinine concentrations using the Abbott enzymatic method. Lithium heparin directly interferes in the Abbott Architect enzymatic creatinine assay. Under-filling lithium heparin tubes can lead to clinically significant over-estimation of creatinine results by this assay. Users of this assay should be aware of the potential for spurious results in small sample volumes collected into lithium heparin tubes and implement robust procedures for identifying and reporting results on these samples.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call